The first experimental step in the validation of an analytical method is the set up for the detection limit (LOD). The
modern methods of limit detection are values statistical determined, which define the degree for an analytical method
can separate a specific chemical substance from the background noise. A common mistake is to considerate the method
detection limit as the lowest concentration which can be measured. Actually, it is the concentration on which we can
decide if an element is present or not ( with an certain trust degree). The detection limits depends on the matrix, the
equipment, the analyst and needs an analytical procedure well-done. Commission decision (2002/657/EC), of 12 august
2002, implementing Council Directive 96/23/EC concerning the performance of analytical methods and the
interpretation of results.